The portable desktop spectrophotometer TS8280 is a spectrophotometer developed by 3NH with independent core spectroscopic technology. It adopts built-in silicon photodiode array (40 groups in double rows) sensors and imported white board, and gives consideration to the speed of measurement and the convenience of operation. The rotating and pressing structure makes the test more convenient. TS8280 portable desktop spectrophotometer color measurement instrument features with repeatability Δ E * ab easily control within 0.05, and inner errors Δ E * ab control within 0.15. This accuracy measurement makes it good used in the lab color accurate analysis and transfer.
Application of TS8280 portable Desktop spectrophotometer:
TS8280 portable desktop spectrophotometer color measurement instrument equipped with Φ30 mm diameter measurement, to adapt to the measurement of large samples, suitable for precise color measurement and quality control in textile and garment printing and dyeing, plastic electronics, ceramics and other industries; and it can be used for fluorescence sample measurement.
Technical Advantages:
1.Adopt international common use d/8 SCI/SCE Synthesis technology
TS8280 spectrocolorimeter adopts D/8(diffused illumination, 8-degree viewing angle) which is widely applicable in the world, and SCI/SCE (specular component included/specular component excluded) Synthesis technology, supporting SCI+SCE simultaneous rapid measurement, and the test time is about 3.2 seconds.
2. Silicon photodiode array sensor (40 groups with double rows)
Double 40 array sensor with larger area, high light saturation, high sensitivity of low light and wide spectral response range ensure the measurement speed, accuracy, stability and consistency of the instrument.
3.A variety of color space, a variety of observation light sources
TS8280 portable desktop spectrophotometer color measurement instrument provides CIE LAB, XYZ, Yxy, LCh, CIE LUV, s - RGB, HunterLab, βxy, DIN Lab99, Munsell (C / 2) color space, and D65, A, C, D50, D55, D75, F1, F2 (CWF), F3, F4 and F5, F6 and F7 (DLF), F8, F9, F10 (TPL5), and F11 (TL84), F12 (TL83 / U30) light sources, which can meet the special measurement requirements under different measurement conditions.
4.Adopt combination full spectrum LED light source and UV light source Each
Full band balanced LED light source ensures sufficient spectral distribution in visible light range, avoids the spectral loss of white LED in specific band, ensures the accuracy of instrument measurement speed and measurement results, and professional UV light source ensures more reliable UV testing.
5.Camera locating can clearly observe the measured area
TS8280 spectrophotometer has a built-in camera for positioning, which can accurately determine whether the measured part of the object is the center of the target through real-time viewing by the camera, thus improving the measurement efficiency and accuracy.
6. Calirbation Certificate
TS8280 spectrophotometer has been verified and tested. After leaving the factory, each instrument is verified according to the measurement standards of authoritative verification departments, and the measurement data are traceable to the National Metrotechnical Institute to ensure the authority of the instrument test data.
7.Industrial-grade HD touch screen, easy to use user interface
Using 7-inch industrial grade hd touch screen, smooth operation, the user interface is easy to use and it makes the operation to become comfortable and convenient.
8.Color management software
SQCX quality management software with TS8280 spectrophotometer is suitable for quality monitoring and color data management in various industries. Data the user's color management, compare color differences, generate test reports, provide multiple color space measurement data, and customize the customer's color management.
9.Optional Accessory
Micro printer, foot switch, rotating bracket can be freely selected according to the needs.
Technical Specification
Model | TS8280 |
Optical Geometry | D/8(diffused illumination, 8-degree viewing angle) |
SCI (specular component included)/SCE (specular component excluded) ; Include UV / excluded UV light source | |
Comply to CIE No.15,GB/T 3978,GB 2893,GB/T 18833,ISO7724-1,ASTM E1164,DIN5033 Teil7 | |
Characteristic | Φ30mm apertures, to adapt to the measurement of large sample; |
Rotary press type structure, more convenient for testing; | |
Accurate measurement, can be used for laboratory color accurate analysis and transmission; | |
Suitable for precise color measurement and quality control in textile and garment printing and dyeing, plastic electronics, ceramics and other industries; It can be used for fluorescence sample measurement. | |
Integrating Sphere Size | Φ152mm |
Light Source | Combined full spectrum LED light source, UV light source |
Spectrophotometric Mode | Flat Grating |
Senso | Silicon photodiode array (double row 40 groups) |
Wavelength Range | 400~700nm |
Wavelength Interval | 10nm |
Semiband Width | 10nm |
Measured Reflectance Range | 0~200% |
Measuring Aperture | Φ30mm/Φ25.4mm |
Specular Component | SCI/SCE |
Color Space | CIE LAB,XYZ,Yxy,LCh,CIE LUV,s-RGB,HunterLab,βxy,DIN Lab99 Munsell(C/2) |
Color Difference Formula | ΔE*ab,ΔE*uv,ΔE*94 ,ΔE*cmc(2:1),ΔE*cmc(1:1) ,ΔE*00, DINΔE99, ΔE(Hunter) |
Other Colorimetric Index | WI(ASTM E313,CIE/ISO,AATCC,Hunter), |
YI(ASTM D1925,ASTM 313), | |
Metamerism Index MI, | |
Staining Fastness, Color Fastness, Color Strength, Opacity | |
8° Glossiness,555 tone classification | |
Observer Angle | 2°/10° |
Illuminant | D65,A,C,D50,D55,D75, F1,F2(CWF),F3,F4,F5,F6, F7(DLF),F8,F9,F10(TPL5), F11(TL84),F12(TL83/U30) |
Displayed Data | Spectrogram/Values, Samples Chromaticity Values, Color Difference Values/Graph, PASS/FAIL Result, Color Simulation, Color Offset |
Measuring Time | About 1.5s (Measure SCI & SCE about 3.2s) |
Repeatability | Chromaticity value: MAV/SCI, within ΔE*ab 0.05 ( When a white calibration plate is measured 30 times at 5 second intervals after white calibration) |
Inter-instrument Error |
MAV/SCI, Within ΔE*ab 0.15 (Average for 12 BCRA Series II color tiles) |
Measurement Mode | Single Measurement, Average Measurement(2-99times) |
Locating Method | Camera Locating |
Dimension | L*W*H=370X240X260mm |
Weight | About 7.8kg |
Battery | AC 24V, 3A Power adapter power supply |
Illuminant Life Span | 5 years, more than 3 million times measurements |
Display | 7-inch TFT color LCD, Capacitive Touch Screen |
Data Port | USB, Bluetooth ®5.0, trigger switch interface |
Data Storage | Standard 1000 Pcs, Sample 30000 Pcs |
Language | Simplified Chinese, English, Traditional Chinese |
Operating Environment | 0~40℃, 0~85%RH (no condensing), Altitude < 2000m |
Storage Environment | -20~50℃, 0~85%RH (no condensing) |
Standard Accessory | Power Adapter, USB Cable, User Guide, PC Software(Download from office website), White and Black Calibration Cavity,Aperture |
Optional Accessory | Micro Printer, Foot Switch, Rotating Bracket |
Notes | Technical parameters are only for reference, subject to the actual sale of the product |
With 12 rigorous quality inspection processes in place, we ensure a high yield rate of up to 98% for batch processed precision parts.
FAQs About Spectrophotometer
There are three measuring apertures for NH310: Φ8mm (standard accessory), Φ4mm (standard accessory), extended Φ8mm (optional accessory)
a. Turning on
b. Removing measuring aperture, install the needed measuring aperture.
c. Select "Settings-Aperture Setting" in main menu to select corresponding measuring aperture.
d. After selecting aperture, the instrument will display "White and Black Calibration" interface. White and black calibration must be performed.
e. Replace completed.
The repeatability is an important parameter for the accuracy of the colorimeter. Threenh Technology portable colorimeter for color measurements tandard deviation ΔE*ab is within 0.03 (Measurement condition: the average value of 30 measurements on the white board at an interval of 3s after calibration). Each colorimeter can pass the qualified inspection before leaving the factory to ensure it conforms to the standard of China Metrology Institute.
a. It’s needed to perform manual calibration for NH310 when first starting.
b. In the subsequent starting, NH310 will perform auto calibration at startup. No need to perform manual calibration.
c. Suggestion: it’s best to perform manual calibration once a week or when NH310 measured data is not accurate.
Yes, we offer free samples for test, you are welcome to order samples to experience our excellent quality and service.
a. If this phenomenon occurred, it means the battery power is not enough. You need to charge the battery.
b. When charge the battery, please insert the battery to the instrument, and plug in the adapter to charge the battery. After 5 minutes, the instrument can be turned on.
c. If the instrument still cannot be turned on after charging 5 minutes, you can pull out the adapter and re-plug in to check it. Or repeat this operation.
d. If the instrument still cannot be turned on after charging 5 minutes, you can turn off the colorimeter and turn on it again to check it. Or repeat this operation.
e. If you have tried all the methods above but the colorimeter still cannot be turned on, please change the battery.