Benchtop Color Haze Meter YH1600
YH1600 Haze Meter can easily realize ASTM D1003 non-compensation method, ISO 13468 compensation method, full transmittance and fog test.
Key Highlights
Double standard ISO & ASTM
CIE LAB, XYZ, Yxy, LCh, s-RGB, βxy color space
Large-size touch control screen
ASTM D1003 uncompensated method, ISO 13468 compensated method, full transmittance, haze test
Independent light source detector and temperature sensor
Measuring and analyzing fog, color and light transmittance software
Color management software
YH1600 Haze Meter can easily realize ASTM D1003 non-compensation method, ISO 13468 compensation method, full transmittance and fog test. Through precise concave grating and 256-pixel CMOS detector, the transmittance curve of transmitted sample can be accurately collected, various chroma data of transmitted sample can be accurately output, and high-precision and repeatable transmittance, fog and chroma data measurement can be realized.
Advantages of YH1600 Haze Meter
1.Double standard ISO & ASTM
Can meet the test standard requirements of different users; meet the standards: ASTM D1003/1044, ISO 13468, ISO 14782, GB/T 2410, JJF 1303-2011, CIE 15.2, JIS K7105, JIS K7361, JIS K 7136
2.Multiple observation light sources
The YH1600 colorimeter provides CIE LAB, XYZ, Yxy, LCh, s-RGB, βxy color space, as well as D65, A, C, D50, D55, D75, F1, F2, F3, F4, F5, F6, F7, F8, F9, F10, F11 and F12 observation light sources, which can meet the special measurement needs under different measurement conditions.
3.Easy operation and faster and more accurate measurement
The YH1600 Haze Meter is equipped with a large-size touch control screen, which is easy and convenient to operate. The 256 pixel dual array CMOS image sensor can realize high accuracy and repeatable transmittance and haze measurement. USB data output unit that interfaces with the laboratory system.
4. Compensation port to make the measurement data more accurate
Meet ASTM D1003 uncompensated method, ISO 13468 compensated method, full transmittance, haze test, provide more accurate test results
5.Dynamic measurement
Independent light source detector and temperature sensor to monitor the light source and environmental changes at all times to ensure reliable test data
6.Convenient measurement and wide sample adaptation
Open measurement area for vertical and horizontal testing to accommodate more samples under test
7.Quality management software
It provides powerful software for measuring and analyzing fog, color and light transmittance, which is suitable for quality monitoring and tabulating management of fog, light transmittance and color data in various industries. Data the management of users on the PC computer, compare the fog, transmittance and color differences, generate test report forms, facilitate customer customization and management.
Application of YH1600 Haze Meter
YH1600 Haze Meter is widely used in glass processing, plastic processing, film, display screen processing, packaging industry, liquid analysis and other aspects.
Product model | YH1600 Haze Meter |
Illumination mode | Transmittance: 0/D (parallel illumination, diffuse receiving) in accordance with standards: ASTM D1003/1044, ISO 13468, ISO 14782, GB/T 2410, JJF 1303-2011, CIE 15.2, JIS K7105, JIS K7361, JIS K 7136 |
Characteristic | The instrument can easily realize ASTM D1003 uncompensated method, ISO 13468 compensation method, full transmittance, fog test. The instrument can accurately collect the transmittance curve of the transmitted sample, and can accurately output various chroma data of the transmitted sample. High hardware configuration, open measurement area, vertical, horizontal testing. It is widely used in glass processing, plastic processing, film, display screen processing, packaging industry, liquid analysis and other aspects. |
Integrating sphere size | Φ 154 mm |
Illumination source | 400 ~ 700 nm combined LED light source (wavelength scalable customized) |
Light splitting mode | Concave grating |
Inductor | 256 pixel dual array CMOS image sensor |
Measuring wavelength range | 400 ~ 700 nm (wavelength scalable customization) |
Wavelength interval | 10 nm |
Half bandwidth | 10 nm |
Determination range of transmittance | 0-100% |
Measuring aperture | Φ20 mm/Φ15 mm/Φ8 mm/Φ4 mm (select single aperture) |
Sample size | Thickness less than 170 mm |
Color space | CIE LAB, XYZ, Yxy, LCh, s-RGB, βxy |
Chromatic formula | ΔEab, ΔE94, ΔEcmc (2:1), ΔEcmc (1:1), ΔE00 |
Other chromaticity indicators | Haze (ASTM D1003/1044, ISO 13468), Transmittance T (ISO), Transmittance T (ASTM) WI (ASTM E313, CIE/ISO, AATCC, Hunter), YI (ASTM D1925, ASTM 313), Absorbance, Cobalt Platinum Index, Gardner Index |
Observer angle | 2 °/10 ° |
Observation light source | D65, A, C, D50, D55, D75, F1, F2, F3, F4, F5, F6, F7, F8, F9, F10, F11, F12 |
Display | Spectrogram, sample chroma value, color difference/graph, color graph, color simulation, pass/fail result |
Measurement time | About 1.5s |
Resolution of fog | 0.01 unit |
Repeatability of nebulization | Φ20 mm aperture, less than 0.08 (after instrument preheating and correction, test the standard nebulizer with the nebulization of about 40 at an interval of 5s for 30 times of standard deviation) |
Inter-stage difference | Φ20 mm aperture, less than 0.4 (after instrument preheating and correction, test the standard deviation between the standard nebulizer of nebulization and the reference value at an interval of 5s) |
Dimensions | Length X Width X Height = 290X211X511 mm |
Weight | Approx. 7.6 kg |
Power supply mode | 24 V DC, supplied by 3A power adapter |
Illumination light source lifetime | > 3 million measurements in 5 years |
Display screen | TFT True Color 7 inch, Capacitive Touchscreen |
Interface | USB, Print Serial , Bluetooth |
Stored Data | 5,000 samples, 20,000 samples |
Language | Simplified Chinese, Traditional Chinese, English |
Operating Temperature Range | 0 to 40 ° C (32 to 104 ° F) |
Storage Temperature Range | - 20 – 50 ° C (-4 – 122 ° F) |
Standard Annex | Power adapter, instruction for use, quality management software (download), data cable, 0% calibration box, measuring caliber |
Optional Accessories | Microprinter, Test Fixture, Standard Fog Disk, Footswitch |
Notes: | Subject to change without notice |
With 12 rigorous quality inspection processes in place, we ensure a high yield rate of up to 98% for batch processed precision parts.
FAQs About Haze Meters
a. Users must specify Standard File Name, Sample File Name and Sample Database File Name when first use.
b.Standard File is use to store standard measurement data. Sample File is used to store sample measurement data. Sample Database File is used to store the measured data exported from sample records.
Threenh Technology with excellent after sale service, Generally, Our Spectrophotometer instrument warranty is one years(Additional support period of 3 years). Within the warranty period, we can offer technique support and parts to parts change, necessarily, our engineers may go to your place for better support. Our instrument have professional After-sales Dept, if you have any technical problem or question with our goods welcome to contact us.
You can choose a colorimters depending on the substrate you are measuring the color of, whether it is opaque, translucent or transparent. In short, you can consider the following factors for choosing the right 3NH colorimeter for your needs: Sample Opacity, Physical Form, Optical Properties needed, Measurement Methods required, Scales and Indices needed, Instrument Standards Conformance (ASTM/ISO/Other), Instrument Geometry, Instrument Performance, Location (laboratory, portability, in-process) and Sample Handling requirements.
By systematically addressing these factors, you can select a colorimeter that optimizes accuracy, efficiency, and value for your specific application.Learn more from our ultimate guide on colorimeter selection. If unsure, consult the 3NH manufacturer for personalized recommendations based on your sample and workflow.
a. It’s needed to perform manual calibration for NH300 when first starting.
b. In the subsequent starting, no need to perform manual calibration.
c. Suggestion: it’s best to perform manual calibration once a week or when NH300 measured data is not accurate.
a. When the colorimeter is disconnected to the PC, "Connection Timeout" will occur. At the moment, you have to check whether the USB cable is well connected the colorimeter and the PC. Then you can re-plug the USB to try out.
b. After ensuring the colorimeter is connected to the PC with USB cable, turn on the colorimeter and enter the main menu. Select "Comm" and press "Enter" key to start communication.
c. Turn off CQCS3 software, and reopen it.